PCT-IH Humidity/Temp Recorder

 

16,000-POINT DATA LOGGER RECORDS BOTH HUMIDITY AND TEMP VS. TIME

 

The Telatemp PCT-IH-DLM is a self-contained data logger module that simultaneously records a trip's relative humidity from 0 to 100% and temperature from -20°C to 60°C with ±.5°C accuracy. The battery operated logger gathers up to 16,000 time and date stamped .3°C accuracy temperature and ±3% RH readings at user selected intervals.

 

START RECORDING. The datalogger can be pre-programmed to start at a particular time or it can be started with a convenient Key Ring Magnetic Starter. Beeping tones sound when the logger is operating. The logger travels with the load to record humidity and temperature exposure at user selected intervals from 1 to 255 minutes or 1 to 4 hours.

 

DATA RECOVERY INTERFACE. At the destination, the trip data can be quickly retrieved by connecting the logger to the Data Recovery Interface and loading the stored information into a PC using the Transit Software. Recorded data is protected against unauthorized use, tampering or duplication with a secure password.

 

VERSATILE TRANSIT SOFTWARE. The menu driven Telatemp software sets up the parameters for temp/time recordings in the Data Logger Module, including the interval between recordings and the cargo's minimum and maximum safe ranges.

 

REPORTING OPTIONS. Recorded data can be displayed and printed in List and Graph modes in °C or °F. A variety of options are available for printing and saving the data. Trip data can be saved to disk in the compact format to be queried at any time or as an ASCII or .SLK file for export to popular spreadsheet, database and word processing programs.

 

PCT-IH HUMIDITY/TEMP SPECIFICATIONS

Data Capacity:

16,000 data points nominal

Temperature Range:

-20°C to +60°C (-4°F to +140°F)

Temperature Accuracy:

±.3°C (±.5°F)

Humidity Range:

0 to 100% RH

Humidity Accuracy:

±3% RH with 0.5% resolution

Timing Accuracy:

±40 sec. per month. Logs in real time

Storage Temperature:

Recommended less than 27°C (80°F)

Power:

3.6V lithium battery; 2 year life

Starting Device:

On-site activation by magnet contact or user programmable for activation at specific date and time

Software:

Menu driven format using Windows

Data Recovery:

Logger mates with Data Recovery Interface for connection to IBM compatible PC

Dimensions:

Diameter: 3.00" (7.6 cm) excluding mounting lug.

Depth: 3.5" (8.9 cm)

Mounting Hole Dia.: 0.32" (0.8 cm)

Unit Weight:

2.8 oz. (80 g)

 

ORDERING INFORMATION

PCT-IH Humidity/Temp Recorder

PCT-IH-DLM

Data Logger Module: One Internal temp sensor and one internal humidity sensor

$293.00

PC-2000-LSP-W

Logger Support Package (W): Includes Data Recovery Interface, Windows Transit Software, Key Ring Magnetic Starter.

$140.00

PC-2000-DRI

Data Recovery Interface: Includes 9-pin serial port connector for IBM and compatibles.

$75.00

PS-2000-TSW-W

Transit Software-Windows

$99.00

PC-2000-KRS

Key Ring Magnetic Starter

$12.50

 

 